| description | Creates Python converter code to transform laboratory instrument data into Allotrope Simple Model (ASM) JSON format. Use when a user asks to convert instrument data to ASM, generate an ASM data instance, write a plate reader converter, map instrument CSV to Allotrope, or validate ASM output. Requires the allotrope-mcp-server MCP server to be connected. |
| license | MIT-0 |
| metadata | {"author":"AWS Labs","version":"1.0.0","mcp-server":"allotrope-mcp-server"} |
Convert Instrument Data to ASM Data Instance
Converts raw laboratory instrument data into a valid Allotrope Simple Model (ASM) JSON data
instance that conforms to the target ASM schema. Covers schema discovery, data parsing, JSON
generation, and schema validation.
ASM Document Structure
ASM organises experimental data as a hierarchy of modular documents. The generated JSON must
mirror this structure exactly.
Technique Aggregate Document ← top-level container (e.g. "plate reader aggregate document")
├── Device System Document ← instrument hardware metadata (model, manufacturer, serial)
├── Data System Document ← software / data-processing metadata (version, file paths)
└── Technique Document[] ← one entry per execution of the technique (e.g. one plate run)
├── Measurement Aggregate Document
│ └── Measurement Document[] ← one entry per individual measurement (e.g. one well)
├── Processed Data Document ← derived results (peak lists, spectra, …)
├── Sample Document ← sample identity and provenance
├── Device Control Document ← acquisition settings and parameters
└── Calculated Data Document ← values computed across measurements or samples
Key points:
- The root JSON key matches the technique aggregate document name from the schema (e.g.
plate reader aggregate document)
$asm.manifest at the root declares the schema — copy this value verbatim from a valid example instance; the URI must end in .manifest
- Technique Document is an array — even a single run must be wrapped in a list
- Measurement Aggregate Document is a container; individual readings live inside its
measurement document array
Parameters
- input_path (required): Path to the instrument data file to convert
- asm_model (required): Name of the ASM model to target (e.g.
plate-reader)
- output_path (optional, default:
<input_path>.asm.json): Destination for the generated ASM JSON
- schema_output_dir (optional, default: current working directory): Directory to save the downloaded schema
Workflow
Step 1: Discover the Target ASM Model
Call the describe_asm MCP tool with the asm_model parameter.
Constraints:
- MUST call
describe_asm before fetching any schema
- MUST record the
asm_json_schema URI for subsequent steps
- SHOULD review
asm_data_instance_examples URIs to understand expected document shape
- If the model name is unclear or
describe_asm errors, call list_asms first to find the correct name
Step 2: Download the ASM Schema
Call fetch_asm_document with the asm_json_schema URI from Step 1.
Constraints:
- MUST pass the
asm_json_schema URI as asm_document_uri
- MUST record the local path returned by the tool for use in Step 7
- SHOULD set
output_dir to a testdata/json-schemas directory to keep schemas organised
Step 3: Inspect a Valid Data Instance
Call fetch_asm_document with one of the asm_data_instance_examples URIs, then read the file.
Constraints:
- MUST download an example before writing any converter code
- MUST identify the
$asm.manifest value, top-level aggregate document key, and measurement document structure
- SHOULD cross-reference the example with the schema to confirm required vs optional fields
Step 4: Parse the Source Data
Read the instrument data file and extract all metadata and measurement values.
Constraints:
- MUST extract: instrument brand, model number, device identifier, and recording timestamp
- MUST parse all well readings into
(row_label, column_label, value) tuples
- MUST normalise the timestamp to ISO 8601 format with UTC timezone offset (
+00:00)
- MUST NOT silently drop records — skip only blank lines and known footer entries (e.g.
Checksum)
Step 5: Write the Converter Script
Implement a Python script that reads the source data and writes a conformant ASM JSON file.
Refer to → CUSTOM-CONVERTER-REQUIREMENTS.md for additional requirements about custom converter files.
Constraints:
- MUST generate a fresh UUID for
ASM file identifier on each run
- MUST map well positions to
<column><row> identifiers (e.g. column A, row 1 → A1)
- SHOULD accept
input_path and output_path as optional CLI arguments with sensible defaults
- SHOULD include Google-style docstrings on all public functions
- MUST NOT hard-code instrument metadata — derive it from the source data header
Step 6: Run the Converter
Execute the script against the source data to produce the ASM JSON output file.
Constraints:
- MUST verify the script exits with code 0
- MUST confirm the output file exists and contains well-formed JSON before proceeding
- MUST confirm the
_map.json field mapping file is also produced alongside the _asm.json file (per CUSTOM-CONVERTER-REQUIREMENTS.md section 6)
Step 7: Validate Against the Schema
Call validate_asm_schema with the generated JSON and the schema path from Step 2.
Constraints:
- MUST pass the output file path as
asm_document_path
- MUST pass the schema path from Step 2 as
asm_schema_path
- MUST confirm
is_valid is true and errors is empty before considering the task complete
- If validation fails, MUST inspect each error, correct the converter, re-run, and re-validate until valid
Step 8: Validate the Field Mapping
Call validate_field_map with the _map.json file produced in Step 6.
Constraints:
- MUST pass the
_map.json file path as field_map_path
- MUST review the
mismatches array in the result; each entry identifies a field whose source and ASM values diverge
- MUST treat a non-empty
mismatches list as a signal to update the converter script, EXCEPT for intentional divergences such as timestamp normalisation (ISO 8601 conversion) and device identifier extraction, which are documented as acceptable in CUSTOM-CONVERTER-REQUIREMENTS.md section 5.5
- SHOULD re-run Steps 6 through 8 after any converter update until all unintentional mismatches are resolved
Examples
Plate Reader CSV → ASM JSON
input_path: tests/testdata/plate_reader_weyland_yutani_instrument_data.csv
asm_model: plate-reader
output_path: tests/testdata/plate_reader_weyland_yutani_instrument_data.asm.json
Expected outcome:
- Schema downloaded to
tests/testdata/json-schemas/...plate-reader.embed.schema
- Converter script written to
convert_plate_reader_csv.py
- Output JSON contains 16 measurement documents (4 columns × 4 rows)
validate_asm_schema returns {"is_valid": true, "errors": []}
validate_field_map reports only the expected intentional mismatches (Instrument serial-number extraction + Recorded timestamp normalisation)
Troubleshooting
Timestamp parse error
Some instruments write timestamps as YYYY-MM-DD:HH:MM:SS (colon between date and time).
Normalise with a regex substitution before calling datetime.fromisoformat.
Well identifier ordering
Source data may lay out columns as headers and rows as the first cell of each data row.
Assemble the well ID as <column_letter><row_number> (e.g. A1), not the reverse.
Schema validation errors on unit
The plate-reader schema requires absorbance values to use unit mAU. Using AU or omitting
the unit field will cause a validation failure.
fetch_asm_document saves to a nested path
The tool mirrors the URI path under output_dir, which can produce a double-nested directory
(e.g. json-schemas/json-schemas/...). Record the exact path returned by the tool and use
that path verbatim in the validate_asm_schema call.