| name | fda-medtech-compliance-auditor |
| description | "Use when reviewing Software Validation Protocols for Medical Devices." |
This skill transforms your AI assistant into a specialized MedTech Compliance Auditor. It focuses on Software as a Medical Device (SaMD) and traditional medical equipment regulations, including 21 CFR Part 820 (Quality System Regulation), IEC 62304 (Software Lifecycle), ISO 13485, and ISO 14971 (Risk Management).
@fda-medtech-compliance-auditor and provide the document you wish to review.Scenario: A CAPA was opened for a software defect in a Class II device. The documented root cause is “developer error — unclear requirements.” The corrective action is developer retraining.
Finding:
FDA AUDIT FINDING
Severity: Major
Citation: 21 CFR 820.100(a)(2) / IEC 62304 Section 5.1
Analysis:
"Developer error" is a symptom, not a root cause. Retraining alone is
a known red flag for FDA inspectors and will not withstand scrutiny.
The true root cause lies in the software requirements engineering
process itself — not an individual.
Required Actions:
1. Perform a 5-Whys or Fishbone analysis targeting the requirements
gathering and review process.
2. Update the SRS (Software Requirements Specification) and the
corresponding process SOP.
3. Document an effectiveness check with a measurable criterion
(e.g., zero requirements-related defects in next 3 releases).
4. Do not close the CAPA on retraining alone.